Chino

IRE-Series - Scanning Type Radiation Thermometers

       The IR-E series is scanning type radiation thermometers using the rotating mirror for measurement of uniaxial temperature patterns without contact and are applicable to the temperature range from 50 to 2000.

IR-E Series

Models
 Thermometer
Distance factors
Models
Elements
50  (50 x 50) IR-E0TF
PbSe (cooling type)
IR-E0TE
PbSe
IR-E0TP
PbS
IR-E0TI
InGaAs
IR-E0TS
Si
100 (100 x 100) IR-E1TF
PbSe (cooling type)
(100 x 50) IR-E1TE
PbSe
IR-E1TP
PbS
IR-E1TI
InGaAs
IR-E1TS
Si
200  (200 x 100) IR-E2TF
PbSe (cooling type)
(200 x 50) IR-E2TE
PbSe
IR-E2TP
PbS
IR-E2TI
InGaAs
IR-E2TS
Si
300  (300 x 60) IR-E3TF
PbSe (cooling type)
IR-E3TE
PbSe
IR-E3TP
PbS
IR-E3TI
InGaAs
IR-E3TS
Si
400  (400 x 70) IR-E4TF
PbSe (cooling type)
IR-E4TE
PbSe
IR-E4TP
PbS
IR-E4TI
InGaAs
IR-E4TS
Si
 Converter
Case types
Signal processing types
Relative value processing
Absolute value processing
Wall mount type
IR-GEW1
IR-GEW2
Panel mount type
IR-GEH1
IR-GEH2

Measuring Ranges and Outputs
Detectors
PbSe
(Cooling type)
PbSe
PbS
InGaAs
Si
Wavelength
(m)
4.0
4.0
2.0
1.55
0.9
Measuring range ()
50 to 300
(50, 100)
200 to 600
(50)
400 to 800
(100, 200)
200 to 300
(50)
600 to 900
(50, 100, 200)
100 to 400
(50, 100)
300 to 800
(50, 100, 200)
500 to 1000
(100, 200, 300)
250 to 400
(50, 100, 200)
700 to 1000
(200, 300, 400)
400 to1000
(50, 100, 200)
300 to 500
(50, 100, 200, 300)
800 to 1200
(200, 300, 400)
500 to 1100
(100, 200, 300)
400 to 700
(50, 100, 200, 300)
900 to 1400
(200, 300, 400)
600 to 1200
(200, 300, 400)
500 to 900
(50, 100, 200, 300)
1000 to 1600
(200, 300, 400)
600 to 1000
(50,100, 200, 300)
1200 to 2000
(200, 300, 400)

Specifications
 Thermometer
Scanning system Optical axis scanning by a rotating mirror
Scanning angle 50 degrees
Scanning speed PbSe, InGaAs, Si: 5, 10, 30, 50 times/second (to be specified)
PbS: 5, 10, 15 times/second (to be specified)
Measuring system Narrow band radiation thermometer
Accuracy rating Less than 400 4
More than 400 1.0% of reading
Measuring distance 0.5 to 10m
Instant view field (mm sq.) Measuring distance/Distance factor
Response time 120s (more than 50 times of scanning speed and 200 of distance factor)
200s (50 times of scanning speed and 50 of distance factor)
2ms (5 times of scanning speed and 50 of distance factor)
For 90% indication
Operating temperature 0 to 50
Power supply 100VAC, 50/60Hz
Power consumption About 40VA
Connection Terminal connection or connector connection
Connection cable length Up to 200m by exclusive cable
Weight About 1kg 
 Converter
Emissivity 0.2 to 1.0
Signal processing Relative value processing (Type 1) IR-GEW1, IR-GEH1
  • Sampling of temperature of any 3 points on the scanning line (P1, P2, P3, Setting dial: 100 equal divisions
  • Maximum value sampling one scanning (P4)
  • Temperature pattern signal
  • Scanning synchronizing signal

  • (signal of 50 degrees scanning)
Absolute value processing (Type 2) IR-GEW2, IR-GEH2
  • Sampling of points at 0 to A mm from left and right edges of sampling object (plate) (P1 - left, P2 - right)
  • Sampling of points at 0 to 100% from left edge of sampling object (plate) (P2 ) Sampling start: From left, Sampling end: Right Setting dial: 100 equal divisions (Measuring distance and A are to be specified.)
  • Maximum value sampling one scanning (P4) 
  • Temperature pattern signal 
  • Scanning synchronizing signal

  • (signal of 50 degrees scanning)
Output signal Sampling signal: 0 to 10mVDC, 4 to 20mADC
Temperature pattern signal: 0 to 5VDC, 4 to 20mADC
Scanning synchronizing signal: TTL level or open collector
Power supply 100, 110, 200 or 220VAC (to be specified), 50/60Hz
Power consumption Maximum 80VA (including thermometer)
Operating temperature range 0 to 60
Weight About 11kg