Chino

IR-M1000 Series - Moisture Meters

       The IR-M1000 Series is reflection system on-line moisture meters utilizing the infrared absorption of moisture. Converting capabilities are built into the compact designed detector unit for easy installation and operation. Maximum 99 calibration curves can be stored into memory for many applications. The unit can be used by itself or by combining with personal computers as an analog output and a communication function are provided with. A setting display unit connectable with maximum 9 detector units can set various kinds of setting items of the detector unit and displays moisture value.

IR-M1000 Series
Features
  • Built-in converting capabilities
    The converting capabilities are built into the small, lightweight and compact detector unit.
  • Measurement with high-speed and high sensitivity
    By high-performance CPU and improved optical components, high speed (output renewal cycle: 28ms) and high sensitive measurement can be performed.
  • Setting display unit
    The setting display unit has a front panel of 96 x 96 mm and can connect with up to 9 sets of the detector unit. The communications interface between the setting display unit and the detector unit is by RS-485.
  • Protective structure conforming to IP-65
    The enclosure of the detector unit and the front panel of the setting display unit conform to IP-65 for severe atmospheric condition.
  • CE-marking
    CE-marking models are available.

Models
 Detector unit
Communications interfaces
Measuring objects
Universal type
High moisture type
Micro moisture type
RS-485 (standard)
IR-M1100S
IR-M1200S
IR-M1300S
RS-232C
IR-M1100R
IR-M1200R
IR-M1300R
RS-422A
IR-M1100A
IR-M1200A
IR-M1300A
 Setting display unit
Models
Communications interfaces
IR-GMEG1S RS-485
IR-GMEG1R RS-232C
IR-GMEG1A RS-422A

Specifications
 Detector Unit
Measuring system Infrared reflection type 3-wavelength system
(2-wavelength system for micro moisture type)
Measuring distance 300 mm (Distance of 200 to 400mm is possible.)
Measuring area About 50 x 50mm (at measuring distance of 300mm)
30 x 30mm for small diameter type
(30 x 30mm/measuring distance 200mm for micro moisture type)
Moisture output 4 to 20mADC (Load resistance: less than 500)
Correction input: 4 to 20mADC (option)
(For correction of sample temperature, measuring distance, etc.)
Communications output RS-485, RS-232C or RS-422A (to be specified)
RS-485 for connection with the setting display unit
Output renewal cycle 28ms
Display LED 5-digit
Calibration curve number - LED 2-digit
Setting By keys or communications
Calibration curve Linear to cubic expression or broken line
Numbers of calibration curve 99 curves
Calibration curve correction Zero and span correction
Calibration curve creation regression calculation
Calibration By a checking plate
Smoothing 0 to 99 seconds
Working temperature range 0 to 50 (Air is necessary for more than 45.)
Rated supply voltage 24VDC
Power consumption Maximum 36VA
Connection Terminals connection
Casing Aluminum casing, drip-proof structure (conforming to IEC529 IP65)
Mounting Suspension system using four M8 bolts
Weight About 4kg
With a power supply unit IR-WEP
 Setting display unit
Detector unit input RS-485, connectable with up to 9 detector units
Analog output 4 to 20mADC (Load resistance: less than 500), 2 outputs
Communication output RS-232C, RS-485 or RS-422A (to be specified)
Output renewal cycle 28ms x (quantity of detector unit)
Display Data - LED 5-digit
Calibration curve number - LED 2-digit
Detector unit number - LED 1-digit
External setting Setting of detector unit number, calibration curve number, calibration, preset output and hold by external contact
Alarm High/low alarms, Contact 2 outputs (H-C-L)
Working ambient temperatur 0 to 50
Rated supply voltage 100 to 240VAC 50/60Hz
Power consumption Maximum 15VA
Casing ABS resin, Front panel - Drip-proof structure (conforming to IP-65)
Weight About 0.6kg
 Power supply unit
Output voltage 24VDC
Output current 2.1A
Working ambient temperature 0 to 50
Supply voltage 100 to 120VAC, 200 to 240VAC, switching type, 47 to 450Hz
Power consumption Maximum 160VA
Weight About 380g
 CE-marking specifications
Analog output

 

4 to 20mADC (Load resistance: less than 500),1 output
(1) Output accuracy - 0.5%of full scale
(2) Stability - 3% of full scale under EMC test environment 
Rated power supply 24VDC, allowable voltage fluctuation - 10% of rated value
Power consumption About 10VA
<Standard>
EN55011 group 1 class A, EN50082-2 (industrial environment)
<Directive> 
89/336/EEC, 92/31/EEC (amendment), 93/68/EEC (amendment)
 Package software IR-WEF1(for IR-M1000, IR-M2000)
Contents

 

Data logging, trend display, data storage and calibration curve creationUp to 9 detector unit s are connectable. An electronic balance is connectable on the creation of calibration curve.
Environment IBM PC/AT compatible
OS Windows 95 Contact input board (option): IBX-2131 (ISA bus) 

Home | About Us | Products | FAQ | What's New | Contact Us

Copyright © 2001, JD Instruments, Inc.
Site designed and maintained by

Web Specialists, Inc.