IR-M2000 Series - Fiber Type Infrared Moisture Meters

       The IR-M2000 Series is fiber type reflection system on-line moisture meters utilizing the infrared absorption of moisture. The sensing part is an optical fiber and can be installed at narrow space and high temperature environment.

       Converting capabilities are built into the compact designed detector unit for easy installation and operation. Maximum 99 calibration curves can be stored into memory for many applications. The unit can be used by itself or by combining with personal computers as an analog output and a communication function are provided with.

       A setting display unit connectable with maximum 9 detector units can set various kinds of setting items of the detector unit and displays moisture value.

IR-M2000 Series
Features
  • High sensitive optical fiber
    The measurement is stable by high sensitive optical fiber. A reflection type optical fiber with lens can measure objects at long measuring distance.
  • Compact sensor part
    The compact and lightweight sensing part using the optical fiber can be used at narrow spaces and places where the direct sighting is difficult.
  • For location in environment at 150
    Any water-cooling in the environment at 150 is not required by the heat-sensitive fiber.
  • Setting display unit
    The setting display unit has a front panel of 96 x 96 mm and can connect with up to 9 sets of the detector unit. The communications interface between the setting display unit and the detector unit is by RS-485.
  • Protective structure conforming to IP-65
    The enclosure of the detector unit and the front panel of the setting display unit conform to IP-65 for severe atmospheric condition.

Models
 Detector unit
Measuring objects
Communications interfaces
Applicable fibers
Universal types
High moisture types
IR-M2100S
IR-M2200S
RS-485 (standard)
Reflection type
IR-M2100R
IR-M2200R
RS-232C
IR-M2100A
IR-M2200A
RS-422A
IR-M2150S
IR-M2250S
RS-485 (standard)
Transparent type
IR-M2150R
IR-M2250R
RS-232C
IR-M2150A
IR-M2250A
RS-422A
 Fiber
Models 
Types 
IR-WCRN  Reflection type 
IR-WCRE  Reflection type with lens 
IR-WCT  Transparent type* 
* For liquid, a measuring cell for liquid is separately necessary.
 Setting Display Unit
Models 
Communications interfaces 
IR-GMEG1S  RS-485 
IR-GMEG1R  RS-232C 
IR-GMEG1A  RS-422A 

Specifications
 Detector Unit
Measuring system Infrared reflection type 3-wavelength system 
Moisture output 4 to 20mADC (Load resistance: less than 500)
Correction input 4 to 20mADC (option)
(For correction of sample temperature, measuring distance, etc.)
Communications output RS-485, RS-232C or RS-422A (to be specified)
RS-485 for connection with the setting display unit
Output renewal cycle 28ms
Display LED 5-digit
Calibration curve number - LED 2-digit
Setting By keys or communications
Calibration curve Linear to cubic expression or broken line
Numbers of calibration curve 99 curves
Calibration curve correction Zero and span correction
Calibration curve creation regression calculation
Calibration By a checking plate
Smoothing 0 to 99 seconds
Working temperature range 0 to 50  (Air is necessary for more than 45.)
Rated supply voltage 24VDC
Power consumption Maximum 36VA
Connection Fiber - Connector connection
Power supply/signal line -Terminal connection
Casing Aluminum casing, drip-proof structure (conforming to IEC529 IP65)
Weight About 4kg
With a power supply unit IR-WEP
 Fiber
Measuring distance and measuring diameter Reflection type - 20mm at 15mm, 50mm at 50mm
Reflection type with lens - 25mm at 25mm, 40mm at 100mm
Fiber length Reflection type, Reflection type with lens - 1.5m (up to 10m)
Transparent type - 2m (up to 20m)
Allowable bending radius Reflection type, Reflection type with lens - R100mm
Transparent type - R250mm
Working temperature range 0 to 150
Fiber protection Stainless flexible tube
Purge air flow Reflection type - 5 to 20Nl/min
Reflection type with lens - Without air purge function
 Setting Display Unit
Detector unit input RS-485, connectable with up to 9 detector units
Analog output 4 to 20mADC (Load resistance: less than 500), 2 outputs
Output renewal cycle 28ms x (quantity of detector unit)
Display Data - LED 5-digit
Calibration curve number - LED 2-digit
Detector unit number - LED 1-digit
External setting Setting of detector unit number, calibration curve number, calibration, preset output and hold by external contact
Alarm High/low alarms, Contact 2 outputs (H-C-L)
Working ambient temperature 0 to 50
Rated supply voltage 100 to 240VAC 50/60Hz
Power consumption Maximum 15VA
Casing ABS resin, Front panel - Drip-proof structure (conforming to IP-65)
Weight About 0.6kg
 Power Supply Unit
Output voltage 24VDC
Output current 2.1A
Working ambient temperature 0 to 50
Supply voltage 100 to 120VAC, 200 to 240VAC, switching type, 47 to 450Hz
Power consumption Maximum 160VA
Weight About 380g
 Package Software Contents
Contents

 

Data logging, trend display, data storage and calibration curve creation
Up to 9 detector unit s are connectable. An electronic balance is connectable on the creation of calibration curve.
Environment IBM PC/AT compatible
OS Windows 95 Contact input board (option): IBX-2131 (ISA bus) 